    Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5; M001314021 IMAGER II/5/ST/65/035 BX…

    Boston Scientific Corporation

    Recall open, classified, 24 August 2026

    FDA record (opens in a new tab)

    What this action means

    A recall is a correction or removal. It may be as small as a label change. This record does not state how many patients were affected or whether harm was caused by the device.

    Reason given

    Tip may fracture or detach, leading to device fragments remaining within the patient or migrating through the vasculature. Patient impact may include prolonged procedure or life-threatening embolism.

    Record

    Number
    Z-3214-2026
    Pathway
    Recall
    Action
    Recall open, classified
    Action date
    24 August 2026
    Applicant or firm
    Boston Scientific Corporation
    Device name
    Imager II Angiographic Catheters M001314001 IMAGER II/5/ST/65/038 BX5; M001314011 IMAGER II/5/ST/100/038 BX5; M001314021 IMAGER II/5/ST/65/035 BX…
    Product code
    DQO
    Review panel
    Cardiovascular
    Device class
    2
    Recall status
    Open, Classified
    Recall event
    99687

    openFDA data are unvalidated and can lag. Use the FDA record as the source of truth. CardioTechX is educational and does not provide regulatory or medical advice.