Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200
Boston Scientific Cork, Ltd.
Recall open, classified, 16 September 2026
What this action means
A recall is a correction or removal. It may be as small as a label change. This record does not state how many patients were affected or whether harm was caused by the device.
Reason given
The product is defective due to insufficient UV curing of the hydrophilic coating which is present on the outer surface of the distal segment of the microcatheters and raised ridges on nitinol sleeves due to laser cutting misalignment.
Record
- Z-0223-2027
- Recall
- Recall open, classified
- 16 September 2026
- Boston Scientific Cork, Ltd.
- Direxion Torqueable Microcatheter, NGMC/SINGLE/021/STRAIGHT/1RO/105; Material Number: M001195200
- KRA
- Cardiovascular
- 2
- Open, Classified
- 99906